A temperature mapping study shows how temperature varies across a storage space, where the hot and cold spots are, and whether every location stays within its required range. It's expected under GMP and GDP for any space storing temperature-sensitive medicines, and it's often the first thing an auditor asks to see. This guide walks through a complete mapping study, from writing the protocol to placing loggers, analysing results and writing the report, with practical tips for warehouses, cold rooms, freezers and fridges.

Key takeaways

  • Mapping finds hot and cold spots and proves a space holds its required temperature everywhere, not just at one sensor.
  • Plan everything in a written protocol before placing a single logger.
  • Place loggers in a 3D grid, plus near doors, cooling units, walls and other risk points.
  • Run the study long enough to capture normal operation, commonly around 7 days, and map warehouses in summer and winter.
  • Use the results to decide where permanent monitoring sensors go.
Contents

What Is a Temperature Mapping Study?

A temperature mapping study is a documented test in which multiple calibrated data loggers record temperature, and often humidity, at many points across a storage space at the same time. The combined data shows how conditions vary in three dimensions and over time.

Mapping answers three questions:

  1. Does every part of the space stay within the required range?
  2. Where are the warmest and coldest points?
  3. Where should permanent monitoring sensors be placed?

It's usually part of the qualification of a storage area or equipment, typically in operational qualification (OQ) for an empty space and performance qualification (PQ) for a loaded space.

Why Temperature Mapping Is Required

Temperature-sensitive products must be stored within their labelled conditions everywhere they sit, not just near a thermostat. In most storage spaces, temperature isn't uniform: air near doors, external walls, lights, cooling units and the top of racks behaves differently from the centre.

Mapping is expected under GMP and GDP guidance, including WHO's technical guidance on temperature mapping of storage areas. Regulators and auditors typically ask for mapping reports for warehouses, cold rooms, freezers and refrigerators used for medicines, vaccines and biologics.

Without mapping, you can't show that products stored in corners, high racks or near doors were safe, and you can't justify where your monitoring sensors are.

When Should You Do a Mapping Study?

  • Before first use of a new storage area or piece of equipment.
  • After significant changes: new racking or layout, HVAC or cooling changes, major repairs, or change of use.
  • Seasonally, for spaces affected by outside conditions, especially warehouses (usually summer and winter).
  • Empty and loaded, since stock changes airflow.
  • Periodically, as defined in your quality system, commonly every one to three years, or after a trend of excursions.

Step 1: Write the Mapping Protocol

The mapping process

  1. Step 1Protocol
  2. Step 2Loggers
  3. Step 3Placement
  4. Step 4Run study
  5. Step 5Analyse
  6. Step 6Report
  7. Step 7Act
The seven steps of a mapping study, covered one by one below.

Never start a mapping study without an approved protocol. It defines what you'll do and how you'll judge the result, before you see any data.

A mapping protocol usually includes:

  • Objective and scope: which space, which conditions (empty, loaded, summer, winter).
  • Description of the space: dimensions, layout drawing, racking, doors, cooling and HVAC units, existing sensors.
  • Acceptance criteria: the required range (for example 2–8°C or 15–25°C), and how excursions will be judged.
  • Equipment: number and type of loggers, accuracy, calibration status and certificate references.
  • Logger positions: a numbered placement plan with a drawing.
  • Logging interval and duration: for example every 1–5 minutes for 7 days.
  • Test conditions: door-opening tests, power-failure or recovery tests, if included.
  • Data analysis method: how you'll find hot and cold spots, calculate statistics and MKT if needed.
  • Responsibilities and approvals: who executes, reviews and approves.
  • Deviation handling: what happens if a logger fails or something unexpected occurs.

Step 2: Choose and Prepare the Data Loggers

The quality of your mapping depends on the loggers. Check that they:

  • Are accurate enough: commonly ±0.2°C to ±0.5°C, depending on the range required. Tight ranges need better accuracy.
  • Have valid calibration: traceable certificates, ideally from an NABL or ISO 17025 accredited lab, covering your temperature range.
  • Cover the range: for freezers and ultra-low freezers, use probe loggers rated for the temperature.
  • Measure humidity if the space has humidity limits.
  • Are time-synchronised: all loggers set from the same computer clock, with the same start time and interval.
  • Have enough memory and battery for the full study at your interval.

Label every logger with its position number, and record each serial number against its position.

See: temperature mapping data loggers · probe data loggers for freezers

Step 3: Plan Logger Placement

Placement is the most important part of the study. Combine a 3D grid with risk-based positions.

The 3D grid

Divide the space into a grid across its length and width, and place loggers at several heights at each grid point, typically low, middle and high. For tall racking, add more heights.

3D logger grid

DoorEvaporatorLowMiddleHighGrid logger (×18)Risk point
Loggers sit on a grid across the floor plan, repeated at low, middle and high levels, plus extra risk points at the door and in front of the evaporator.

Risk-based positions

Add loggers wherever conditions are likely to differ:

  • Near doors and loading docks
  • Near cooling units, evaporators and air outlets
  • Near external walls, windows and roof
  • Near lights and other heat sources
  • In corners and dead air zones
  • Next to existing monitoring sensors (to compare readings)

Typical number of loggers

SpaceTypical logger countPlacement notes
Fridge or freezer (single door)About 9–15Corners and centre of each shelf level, plus near the door
Walk-in cold roomAbout 15–303D grid plus door, evaporator and corners
Small warehouseAbout 20–40Grid at 3 heights plus doors and walls
Large warehouse50 or moreGrid across floor area at several rack heights, plus docks, HVAC and roof

These are practical ranges. Your protocol should justify the final number based on the size, layout and risk of the space. In warehouses, grid spacing of a few metres is common, with closer spacing near risk areas.

Placement tips

  • Don't place loggers directly against walls, cooling coils or in direct airflow unless that's what you're testing.
  • Keep the logger in the space where product is stored, at product height.
  • For loaded studies, place loggers among the product, not on top of cartons.
  • Photograph every logger in position for the report.

Step 4: Run the Study

  • Duration: long enough to capture normal daily and weekly cycles, including weekends, commonly around 7 days. Equipment like fridges may be shorter, around 24–72 hours, if justified in the protocol.
  • Logging interval: commonly 1 to 5 minutes, so short spikes aren't missed.
  • Operate normally: for performance mapping, the space should run as it does every day, including deliveries, door openings and defrost cycles.
  • Record events: keep a log of door openings, deliveries, defrosts, power cuts and anything unusual, with times. This explains spikes later.

Optional challenge tests

Many studies add:

  • Door-opening test: open the door for a set time and record how fast the temperature recovers.
  • Power-failure test: switch off power and record how long the space stays within range, then how fast it recovers. This shows how much time you have to respond to an alarm.

Step 5: Analyse the Data

After the study, download every logger and combine the data. For each position, calculate:

  • Minimum, maximum and average temperature
  • Time outside the acceptance range, if any
  • Mean kinetic temperature (MKT), where your protocol or quality system requires it

Then for the whole space:

  • Identify the hot spot and cold spot: the positions with the highest and lowest readings, and when they occurred.
  • Check uniformity: the spread between the warmest and coldest positions.
  • Review events: match spikes with door openings, defrosts or other logged events.
  • Assess challenge tests: recovery times after door opening and power failure.

A temperature-over-time chart for all positions, plus a colour-coded layout drawing (a "heat map") showing average or maximum readings at each position, makes results easy to understand.

Example: 7-day mapping data

0°C2°C4°C6°C8°C10°CDay 0Day 1Day 2Day 3Day 4Day 5Day 6Day 7Door held openLimit 8°CLimit 2°CHot spotCold spotCentreMid-rack
Each line is one logger position. The warmest position briefly approaches 8°C when a door is held open; the coldest sits near 2.5°C by the evaporator.

Example: heat map of average readings

DoorEvaporator6.8°6.1°5.6°5.9°7.2°5.9°5.0°4.6°4.9°6.3°5.2°4.3°3.1°4.0°5.5°Average °C7.26.25.34.53.0Hot spot 7.2°Cold spot 3.1°
Interpolating between logger positions shows where the room runs warm and cold. Here the warmest corner is by the door, the coldest point is in front of the evaporator.

Step 6: Write the Mapping Report

A mapping report normally includes:

  • Summary and conclusion: pass or fail against the acceptance criteria
  • Reference to the approved protocol
  • Description of the space and test conditions
  • Logger list with serial numbers, positions and calibration references
  • Placement drawing and photos
  • Results tables per position and charts
  • Hot spot and cold spot identification
  • Challenge test results
  • Deviations and their impact
  • Recommendations: permanent sensor positions, unusable zones, corrective actions
  • Review and approval signatures

Step 7: Act on the Results

  • Place permanent monitoring sensors at the hot and cold spots, and at points representative of the stored product.
  • Mark no-storage zones where conditions fall outside the range, for example directly in front of a cooling unit or near a dock door.
  • Fix problems, such as poor airflow, gaps in insulation or faulty door seals, and re-map if needed.
  • Set alarm limits and delays based on recovery times from the challenge tests.

Continuous monitoring then watches the critical points found during mapping. See wireless temperature data loggers and multi-point monitoring systems.

Mapping Different Spaces

Warehouses

Large, tall spaces with the biggest variation. Map in summer and winter, include high racking positions, and pay extra attention to roofs, external walls, dock doors and HVAC outlets. Humidity mapping is often required too. See temperature and humidity data loggers.

Cold rooms

Watch the evaporator area, door and corners. Include defrost cycles in the study, and run door-opening and power-failure tests to set alarm delays.

Refrigerators and freezers

Map each shelf level, including near the door and the back wall. For freezers at -20°C or -80°C, use probe loggers so the electronics aren't inside the freezer.

Reefer vehicles and containers

Map the cargo space front, middle and rear, at floor, middle and top, near the doors and the refrigeration unit, during a representative journey or a static test.

Common Temperature Mapping Mistakes

  • Too few loggers, or placing them only where it's convenient.
  • Missing risk points like doors, cooling units and high racks.
  • Loggers not time-synchronised, making position-to-position comparison unreliable.
  • Expired or missing calibration certificates.
  • Studies that are too short to capture weekends, defrosts or normal operations.
  • Mapping only once, without seasonal or post-change studies.
  • Not acting on the results: no permanent sensors at hot or cold spots.

Temperature Mapping Checklist

Before the study

  • Protocol written and approved
  • Acceptance criteria defined
  • Loggers calibrated, certificates on file
  • Loggers time-synchronised and configured with the same interval
  • Placement plan and drawing prepared
  • Loggers labelled with position numbers

During the study

  • Loggers placed and photographed
  • Normal operations maintained
  • Event log kept (doors, deliveries, defrosts, power)
  • Challenge tests performed, if included

After the study

  • All loggers downloaded and data combined
  • Min, max, average, time out of range and MKT calculated
  • Hot and cold spots identified
  • Report written, reviewed and approved
  • Permanent sensor positions and no-storage zones defined
  • Next mapping date scheduled

Download checklist (PDF)

Frequently Asked Questions

What is a temperature mapping study?

A temperature mapping study is a documented test in which calibrated data loggers record temperature, and often humidity, at many points in a storage space at the same time. It identifies hot and cold spots and shows whether the whole space stays within its required range.

How long should a temperature mapping study last?

Most studies for warehouses and cold rooms run for around 7 days, to capture normal daily and weekly operations, including weekends. Smaller equipment like fridges may be mapped over a shorter period if justified in the protocol.

How many data loggers are needed for temperature mapping?

It depends on the size and layout of the space. A single fridge or freezer typically needs around 9 to 15 loggers, a walk-in cold room 15 to 30, and a large warehouse 50 or more, placed in a 3D grid plus risk points.

How often should temperature mapping be done?

Map before first use, after significant changes to the space or equipment, seasonally for warehouses, and periodically as defined by your quality system, commonly every one to three years.

What is the difference between empty and loaded mapping?

Empty mapping shows how the space performs without product and is usually part of operational qualification. Loaded mapping, done with product in place, shows real-world performance and is usually part of performance qualification, since stock changes airflow.

What should a temperature mapping report include?

It should include the protocol reference, space description, logger list with calibration references, placement drawings and photos, results per position, hot and cold spots, challenge test results, deviations, conclusions and recommendations, with approval signatures.

Can we do temperature mapping in-house?

Yes, if you have suitable calibrated loggers, trained staff and a documented protocol. Many companies map in-house for routine and seasonal studies and use a specialist service for complex or first-time qualifications.

Related loggers

Related guides

Planning a mapping study?

Tell us the space you need to map and how many positions you need, and we'll recommend calibrated loggers and send a quote.